Cement fineness negative pressure sieve analyzer professional manufacturer

Beijing Zhongke Luda Test Instrument Co., Ltd. is a professional manufacturer of cement fineness negative pressure sieve analyzer . It has strong technical force and sophisticated equipment, and can provide users with high quality, high performance products and perfect technical services.

Cement fineness negative pressure sieve analyzer

Model: FYS-150B type

Cement fineness negative pressure sieve analyzer cement negative pressure sieve analyzer according to the national standard CB1345-91 (cement fineness inspection method 80uM sieve analysis method) for the inspection of the special instrument with simple structure, easy operation, reduce energy consumption, etc., is the test Essential equipment for laboratory and cement professional research units.

Cement fineness negative pressure sieve analyzer cement negative pressure sieve analyzer

The main technical parameters:

1, sieve analysis test fineness: 80um
2, sieve analysis automatic control time: 2min
3, working negative pressure adjustable: 4000-6000pa
4, working voltage: 220V

Cement fineness negative pressure sieve analyzer operating procedures (1) Place the cement fineness negative pressure sieve analyzer on the level ground, open the back door to check whether all parts are properly installed.
(2) Weigh the cement sample 25g, pour it into the sample sieve and cover the sieve cover, then start the instrument. After the machine is stopped, remove the sample sieve, pour the remaining objects into the balance and weigh the results. .
(3) The working negative pressure should be adjusted within the range of document.write('') -document.write('') 4000~-6000Pa. The instrument automatically stops within 2 minutes. (4) After each use, apply a brush to gently brush both sides of the test screen, and put the screen against the light. After checking the screen hole, save the test sieve in a dry container or plastic bag. When the test sieve is clogged, it can be reversed on the sieve seat, covered with a sieve cover to suck back, and then brushed with a brush. If the screen is severely blocked, the test sieve can be immersed in water for a while and then brushed.
(5) If the cement fine powder in the dust collection bottle is almost full, the dust collection bottle should be pulled out from the cyclone cylinder and then refilled.
(6) The vacuum cleaner should pay attention to regular cleaning to keep the dust bag clean.


The cement fineness negative pressure sieve analyzer is the main product of Beijing Zhongke Luda Test Instrument Co., Ltd., with reliable quality and superior performance. It is widely used in various quality inspection centers, universities and colleges, construction units, and welcome to buy Beijing Zhongke Road. Cement Fineness Negative Pressure Screener

:document.write('') document.write('') : : Zhao Yang

Facial Treatments

Fei County Mingtao Electronic Co.,Ltd , https://www.yourbeautyassistant.com